Search GimmeJob

Type at least 2 characters. Enter also searches by meaning.
QA interview question
Regulated domainsLeadSpecialistScenario

What should regulated change control require for a validated data migration after a safety-impacting defect?

Answer

Open a controlled change linked to the defect, affected requirements, risk assessment and intended migration, with independent review appropriate to the product's safety impact. Approve versioned transformation rules and reconciliation criteria before execution, qualify tools where required and preserve source, exceptions, approvals and results in an audit trail. Test rollback, partial failure, data integrity and representative safety workflows, then complete impact-based regression and formal release authorization before closing the change.

Strong answer includes

  • links the change to risk, requirements and independent approval
  • preserves migration rules, exceptions and reconciliation evidence
  • covers rollback, integrity and safety-impacting regression

Practical example

For example, after a defect causes a small number of infusion-pump dosage records to migrate with a unit-conversion error, the change-control record links the change to the original defect ticket, a documented risk assessment classifying it as safety-impacting, and independent QA sign-off separate from the developer who wrote the fix. The team approves the corrected transformation rule and reconciliation query in advance, executes the re-migration in a validated environment, tests a rollback to the pre-migration snapshot, and only closes the change after a quality lead formally authorizes release with the full audit trail attached.

#regulated#change-control#migration#safety

Sources

Computer Software Assurance for Production and Quality Management System Software (final guidance, February 2026)US FDA · Risk-based assurance and objective evidence for quality-system softwareFDA recognition of IEC 62304US FDA · Medical-device software lifecycle processes, safety classification and maintenanceEudraLex Volume 4, Annex 11 — Computerised SystemsEuropean Commission · GMP computerized-system validation, data integrity, change and continuity controls